ORIGINAL ARTICLE
Year : 2016  |  Volume : 15  |  Issue : 2  |  Page : 91-95

Effect of radionuclide activity concentration on PET-CT image uniformity


1 Department of Nuclear Medicine, CM Johannesburg Academic Hospital, University of the Witwatersrand, Johannesburg, South Africa; Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana; Medical Radiation Physics Centre, Radiological and Medical Sciences Research Institute, Ghana Atomic Energy Commission, Accra, Ghana
2 Department of Nuclear Medicine, CM Johannesburg Academic Hospital, University of the Witwatersrand, Johannesburg, South Africa
3 Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana, Accra, Ghana

Date of Web Publication3-Mar-2016

Correspondence Address:
Francis Hasford
Medical Radiation Physics Centre, Radiological and Medical Sciences Research Institute, Ghana Atomic Energy Commission, P.O. Box LG 80, Legon, Accra

Login to access the Email id


DOI: 10.4103/1450-1147.167578

PMID: 27134558

Rights and Permissions
   Abstract 

Assessment of radionuclide activity concentration on positron emission tomography-computedr tomography (PET-CT) image uniformity has been carried out quantitatively. Tomographic PET-CT images of cylindrical phantom containing F-18 fluorodeoxyglucose (FDG) activity concentration was acquired and used for the assessment. Activity concentrations were varied and PET-CT images were acquired at the constant acquisition parameters of time, matrix size, and reconstruction algorithm, respectively. Using midtransaxial image slices, quantitative index of nonuniformity (NU), and coefficient of uniformity variation were estimated for the different activity concentrations. Maximum NUs of 17.6%, 26.3%, 32.7%, 36.2%, and 38.5% were estimated for activity concentrations of 16.87 kBq/mL, 14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL, respectively. The coefficient of uniformity variation established an inverse quadratic relationship with activity concentration. Activity concentrations of 16.87 kBq/mL, 14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL produced uniformity variations of 1.47%, 2.52%, 4.23%, 5.12%, and 4.98%, respectively. Increasing activity concentration resulted in decreasing coefficient of uniformity and hence, an increase in image uniformity. The uniformity estimates compared well with the standards set internationally.

Keywords: Activity concentration, counts of activity, image uniformity, positron emission tomography-computed tomography, radionuclide


How to cite this article:
Hasford F, Wyk BV, Mabhengu T, Vangu MD, Kyere AK, Amuasi JH. Effect of radionuclide activity concentration on PET-CT image uniformity. World J Nucl Med 2016;15:91-5

How to cite this URL:
Hasford F, Wyk BV, Mabhengu T, Vangu MD, Kyere AK, Amuasi JH. Effect of radionuclide activity concentration on PET-CT image uniformity. World J Nucl Med [serial online] 2016 [cited 2020 Jan 23];15:91-5. Available from: http://www.wjnm.org/text.asp?2016/15/2/91/167578


   Introduction Top


The acquisition of accurately coregistered anatomic and functional images is a major strength of the positron emission tomography-computed tomography (PET-CT) integrated system. However, the additional advantage of this approach is the possibility to use the CT images for attenuation correction of the PET emission data, eliminating the need for a separate, lengthy PET transmission scan. The use of the CT scan for attenuation correction not only reduces whole-body scan by 30-40% but also provides essentially noiseless attenuation correction factors (ACFs) compared to those from standard PET transmission measurements. [1],[2],[3],[4] Since the attenuation values are energy-dependent, the correction factors derived from a CT scan at mean photon energy of 70 keV must be scaled to the PET energy of 511 keV. [5],[6] Integrated PET-CT system has become an established diagnostic modality that is extensively used in the oncology for tumor diagnosis, staging, and radiotherapy treatment planning and monitoring in cardiology for myocardial viability and perfusion, and in neurology for perfusion and neuroreceptor imaging. [7],[8]

Most image distortions in PET-CT images can be associated with distortions of the CT images that propagate into the PET images through the use of CT-based attenuation correction. Similarly, pitfalls arise from the limitations of the standard approach to CT-based attenuation correction that does not account for the presence of artificial implants or CT contrast agents. [3] Furthermore, misalignment effects between the CT and PET acquisitions may create biases in the reconstructed PET images. An advantage of PET imaging is its ability to accurately quantify the distribution of the injected tracer in the body noninvasively. This accuracy relies on several corrections during the image acquisition and reconstruction process but also depends on the accurate knowledge of the amount of injected activity. One of the most widely used metrics in PET imaging is the standard uptake value (SUV) that is defined as:



Physicists are recommended to conduct a qualitative and quantitative review of PET-CT image uniformity during acceptance testing and at least once in every quarter. [9],[10] Qualitative review includes evaluation of the uniformity and noise characteristics of the images and quantitative review includes evaluation of the accuracy of the SUV calibration and the axial variation of the average SUVs. The National Cancer Institute's (NCI) manual of procedures for PET and PET-CT systems recommends an acceptance criterion of less than 10% for axial variation in PET-CT image uniformity. [8] The International Atomic Energy Agency's Human Health Series (HHS) publication on quality assurance for PET and PET-CT systems (HHS No. 1) also recommends a nonuniformity (NU) estimate of less than 1.05 × reference NU. [9] This study evaluates the effect of activity concentration on the uniformity of PET-CT images.


   Materials and Methods Top


A PET-CT integrated system (Biograph 40, Siemens, Memphis, Tennessee, USA) was used in acquiring images for the study. The system has a 40-slice CT component that provides a 360° full scan of an average patient in approximately 40 s. The PET component uses lutetium oxyorthosilicate (LSO) detector that produces fast scintillation, high light yield, and fast light decay time. Dimensions of the LSO detectors are 4 × 4 × 20 mm 3 , giving it a high resolution capability. The system has a 70-cm spacious gantry opening for easy and flexible patient positioning, enhanced patient comfort, and convenience. Cylindrical phantom (Siemens, Memphis, Tennessee, USA) was used for the PET-CT image uniformity assessment. The phantom has internal diameter of 190 mm and a volume of 6.58 × 10 6 mm 3 . The cylinder comes with a lid that has three fill ports through which water or activity is introduced into the phantom.

The cylindrical phantom was filled with water such that it was about three quarters full and 111 MBq F-18 fluorodeoxyglucose (FDG) activity was injected into it. The cylinder was covered with the lid and the phantom was repeatedly inverted to thoroughly mix the activity with the water. More water was added to the cylinder until it was almost completely full. It was repeatedly inverted again to achieve uniform distribution of activity in the phantom. With the phantom lid still closed, the water was injected gradually through the fill port into the phantom using a syringe until the phantom was completely full. The resultant activity concentration was estimated to be 16.87 kBq/mL. Air bubbles were removed by gently shaking the phantom.

The phantom was mounted to the patient bed such that its long axis was parallel to the axis of the scanner and its positioning was done using lasers, as shown in [Figure 1]. Scout view of the phantom was first taken with the CT scanner and used to define the scan boundaries of the phantom. The scan length on the PET system was a two-bed position with the phantom centered in the axial extent of the combined two-bed positions. The CT and PET phantom scans were acquired using standard clinical protocols for body PET-CT imaging in accordance with the manufacturer's recommendation. The PET scan was performed at 4 min per bed position.
Figure 1: Cylindrical phantom under scanning

Click here to view


Axial PET-CT images of the phantom were acquired with the system at 3-mm slice thicknesses. Slices corresponding to the uniform long active part of the phantom were reconstructed with all the corrections applied using the system software. Calculated attenuation correction was applied using the corresponding CT images to avoid noise propagation. Reconstructed transaxial and sagittal slices of the images were displayed and carefully inspected visually for artifacts. One midtransaxial slice was selected for the PET-CT image uniformity assessment by estimating its quantitative index of nonuniformity NU and the coefficient of uniformity variation. The entire study was repeated at reduced F-18 FDG activity concentrations of 14.06 kBq/mL (injection of 92.5 MBq activity), 11.25 kBq/mL (injection of 74 MBq activity), 8.43 kBq/mL (injection of 55.5 MBq activity), and 5.62 kBq/mL (injection of 37 MBq activity) and their corresponding nonuniformity NU indices and coefficients of variation were estimated. The estimated uniformity indices and coefficients of variation for the five activity concentrations were then compared to assess the effect of radionuclide activity concentration on PET-CT image uniformity.

For each activity concentration, one midtransaxial image slice was chosen for uniformity analysis on the PET-CT computer system. Quantitative index of nonuniformity NU and coefficient of variation were estimated by drawing a circular area of 175-mm diameter centered inside the selected image slice. As shown in [Figure 2], a total of 177 square regions of interest measuring 10 × 10 mm 2 were drawn on each slice inside the circular area. For each region of interest (ROI) k in image slice i, the maximum, average, and minimum counts of activity inside the kth ROI were identified and recorded.
Figure 2: PET-CT image slice of phantom (with 8.43 kBq/mL F-18 FDG)

Click here to view


The nonuniformities (NUs) in individual ROIs across the image slice were estimated using the NU expression in Equation 2. [10] The maximum NU among the 177 ROIs in each selected image slice was identified. Coefficient of uniformity variation (CV) across the image slice was also estimated using Equation 3. [10]





where, standard deviation ,



and denote the number of the square ROIs inside the 175 mm diameter circle.


   Results Top


[Table 1] presents NU estimates in the ROIs for the five activity concentrations and [Table 2] presents the coefficients of uniformity variation in the concentrations.
Table 1: Nonuniformities in ROIs for different activity concentrations


Click here to view
Table 2: Coefficient of uniformity variation


Click here to view


A variation of the estimated image uniformity with activity concentration is graphically presented in [Figure 3].
Figure 3: Variation of image uniformity with activity concentration

Click here to view



   Discussion Top


NUs in the 177 ROIs for the five activity concentrations are presented in [Table 1]. The NUs were estimated to assess the PET-CT system's response to homogeneous activity distribution on the micro level. The higher the NU value in a ROI, the less uniform the activity concentration in that particular region. From the study, highest and least NU estimates were attained in image slices having activity concentrations of 5.62 kBq/mL and 16.87 kBq/Ml, respectively. The study observed a pattern where the maximum NU in ROIs increases with decreasing activity concentration. Activity concentrations of 16.87 kBq/mL, 14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL produced maximum NUs of 17.6%, 26.3%, 32.7%, 36.2%, and 38.5%, respectively. The NU estimates, however, represent only single image slices and hence, may not be generalized for the entire volume of phantom.

Given that PET imaging for different activity concentrations were all carried out in equal time durations of 4 min per bed position, F-18 FDG with higher activity concentration was seen to accumulate more counts of activity compared with less activity concentrations. Accumulation of more counts of activity produces more uniform images and hence, NUs in smaller regions of interest are also lower compared with lesser activity concentrations. Other acquisition parameters such as matrix size and reconstruction algorithms remained constant throughout the study. Average counts of activity per ROI for the 16.87 kBq/mL,

14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL activity concentrations were estimated to be 10341.1 cts, 488.3 cts, 280.2 cts, 105.8 cts, and 75.1 cts, respectively.

Coefficients of uniformity variation for the different activity concentrations were estimated to assess the degree of uniformity over the entire surface of the transaxial image slices. The lesser the coefficient of variation, the better the image uniformity and vice versa. The coefficients are estimated in [Table 2] and graphically presented in [Figure 3]. From the results, an inverse quadratic relationship is established with the equation y = 0.0009x 4 -0.0320x 3 + 0.3749x 2 -1.6418x + 7.2007 at R 2 = 1. At constant acquisition factors, increasing activity concentration results in decreasing coefficient of uniformity and hence, an increase in image uniformity. SD estimates, however, showed a linear trend with increasing activity concentration.

Results from the uniformity estimates compare favorably well with NCI's recommendation of an acceptance criterion of less than 10% for axial variation in PET-CT image uniformity. [8] Due to unavailability of reference NU estimates for comparison, acceptance criterion recommended by HHS No. 1 [9] could not be assessed. However, with quantitative estimates of uniformity ranging 1.47-4.98% recorded for the different activity concentrations, it could be implied that the acceptance criterion of HHS No. 1 could also be within range. Qualitative assessment of the image slices from different activity concentrations showed relatively better uniformity in the higher activity concentrations than in the lower concentrations.


   Conclusion Top


As part of acceptance testing and quarterly quality control checks of PET-CT systems, PET-CT image uniformity is carried out to assess the system's response to homogeneous activity distribution. This study has quantitatively assessed the level of uniformity in PET-CT image slices that would otherwise have been difficult to accurately predict qualitatively. The study has demonstrated that PET-CT image uniformity is improved with increasing activity concentration at constant acquisition parameters.

 
   References Top

1.
Shreve P, Townsend DW. Clinical PET-CT in Radiology-Integrated Imaging in Oncology. New York, USA: Springer; 2011. p. 26-7, 40-3, 47-8.  Back to cited text no. 1
    
2.
Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys 1998;25:2046-53.  Back to cited text no. 2
    
3.
Kinahan P, Hasegawa B, Beyer T. X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 2003;33:166-79.  Back to cited text no. 3
    
4.
Von Schulthess GK. Cost considerations regarding an integrated CT-PET system. Eur Radiol 2000;10(Suppl 3):S377-80.  Back to cited text no. 4
    
5.
LaCroix KJ, Tsui BMW, Hasegawa BH, Brown JK. Investigation of the use of X-ray CT images for attenuation compensation in SPECT. IEEE Trans Nucl Sci 1994;41:2793-9.  Back to cited text no. 5
    
6.
Tang HR, Brown JK, Da Silva AJ, Mathay KK, Price DC, Huberty JP, et al. Implementation of a combined X-ray CT-scintillation camera imaging system for localizing and measuring radionuclide uptake: Experiments in phantoms and patients. IEEE Trans Nucl Sci 1999;46:551-7.  Back to cited text no. 6
    
7.
Jadvar H, Parker JA. Clinical PET and PET-CT. New York, USA: Springer; 2005. p. 69-267.  Back to cited text no. 7
    
8.
Boellaard R, O'Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, et al. FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: Version 1.0. Eur J Nucl Med Mol Imaging 2010;37:181-200.  Back to cited text no. 8
    
9.
American College Radiology Imaging Network. PET-PET/CT Technical Procedures: NCI CQIE Manual of Procedures, Ver 3.2. 1: American College Radiology (ACR); 2013. p. 7-9.  Back to cited text no. 9
    
10.
International Atomic Energy Agency. Quality Assurance for PET and PET-CT Systems: Human Health Series No. 1. Vienna, Austria: International Atomic Energy Agency (IAEA); 2009. p. 125-31.  Back to cited text no. 10
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2206    
    Printed89    
    Emailed0    
    PDF Downloaded312    
    Comments [Add]    

Recommend this journal