CASE REPORT
Year : 2015  |  Volume : 14  |  Issue : 1  |  Page : 63-65

Positron Emission Mammography Imaging with Low Activity Fluorodeoxyglucose and Novel Utilization in Core-needle Biopsy Sampling


Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, Tx 75390, USA

Date of Web Publication2-Feb-2015

Correspondence Address:
Stephen Seiler
Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, Tx 75390
USA
Login to access the Email id


DOI: 10.4103/1450-1147.150557

PMID: 25709550

Rights and Permissions
   Abstract 

Positron emission mammography (PEM), a relatively novel breast imaging modality, provides certain advantages over magnetic resonance imaging, including the ability to image biopsy samples. However, the radiation activity associated with PEM has remained a concern in clinical practice. We present a case of an invasive ductal carcinoma that was adequately imaged with a much lower than the standard 185 to 370 MBq activity of 18 F-fluorodeoxyglucose. In addition, we demonstrate ultrasound-guided core-needle biopsy sample imaging with PEM to assess adequacy of sampling, a strategy that has previously only been documented with vacuum-assisted biopsy samples.

Keywords: Activity, biopsy, cancer, fluorodeoxyglucose, positron emission mammography


How to cite this article:
Choudhery S, Seiler S. Positron Emission Mammography Imaging with Low Activity Fluorodeoxyglucose and Novel Utilization in Core-needle Biopsy Sampling. World J Nucl Med 2015;14:63-5

How to cite this URL:
Choudhery S, Seiler S. Positron Emission Mammography Imaging with Low Activity Fluorodeoxyglucose and Novel Utilization in Core-needle Biopsy Sampling. World J Nucl Med [serial online] 2015 [cited 2019 Dec 7];14:63-5. Available from: http://www.wjnm.org/text.asp?2015/14/1/63/150557


   Introduction Top


Positron emission mammography (PEM) is being increasingly utilized in breast cancer imaging. The sensitivity of PEM is comparable to that of breast magnetic resonance imaging (MRI), and its specificity has been shown to be higher than breast MRI in presurgical planning. [1] PEM provides better spatial resolution and increased sensitivity in comparison to whole-body positron emission tomography (PET) imaging. [2],[3] Another advantage PEM renders over whole-body PET is the ability to use a lower activity of 18 F-fluorodeoxyglucose ( 18 F-FDG). Currently, the recommended activity for PEM is 185 to 370 MBq. We present a case of a woman with invasive ductal carcinoma whose malignancy was imaged with PEM with a much lower than the conventionally used activity of 18 F-FDG. In addition, we demonstrate a new application of PEM in confirming adequate sampling of ultrasound-guided core-needle biopsy samples.


   Case Report Top


A 67-year-old female presented to an outside hospital with a right axillary mass [Figure 1]a and b. An ultrasound-guided biopsy confirmed invasive ductal carcinoma. A whole-body PET-computed tomography (CT) scan performed thereafter showed a 1.7-cm mass with an SUV maximum of 5.9 in the right axilla, consistent with the patient's known malignancy, and an adjacent FDG-avid right axillary lymph node [Figure 2].
Figure 1: (a) Spot compression mammographic image of the right axilla of our patient shows a round hyperdense mass. (b) Ultrasound image of the right axilla shows an irregular, hypoechoic mass. On biopsy, this was found to be an invasive ductal carcinoma

Click here to view
Figure 2: Fused coronal PET-CT image of our patient before the axillary lymph node biopsy shows a lobulated FDG-avid mass (star) in the right axilla, which represents the patient's known axillary invasive ductal carcinoma. An adjacent small FDG-avid round structure (arrow) was suspicious for metastatic axillary lymphadenopathy

Click here to view


A repeat whole body PET-CT scan was performed at our institution to assess extent of disease. The patient was given 362.6 MBq of 18 F-FDG at 07:45 on the day of her whole body PET-CT. The scan was performed at approximately 09:18 and demonstrated stable findings in the right axilla. At 13:35, an ultrasound-guided core-needle biopsy of the patient's known FDG-avid right axillary lymph node was performed. PEM imaging (using PEM Flex Solo II, Naviscan) of the right axilla was then performed at 13:47 without administering more 18 F-FDG. PEM imaging relied upon the remainder of the dose of 18 F-FDG that the patient had received six hours earlier [Figure 3]a. These post-biopsy PEM images successfully and adequately displayed the patient's known right breast carcinoma utilizing much lower activity than conventional PEM imaging. In addition, these images showed that the FDG-avid right axillary lymph node had decreased in conspicuity after the biopsy in comparison to the pre-biopsy whole-body PET-CT, an indirect indication of successful biopsy sampling [Figure 3]b. Following PEM imaging of the right axilla, at approximately 14:50, the core biopsy samples were also examined with PEM imaging [Figure 4]a and b, confirming activity within portions of the core biopsy samples.
Figure 3: (a) PEM imaging of the right axillary tail shows the FDG-avid mass in the right axilla, which represents the patient's known invasive ductal carcinoma (IDC). Even with an approximate 37 MBq activity of 18F-FDG in the patient's system, the patient's carcinoma is well-visualized. (b) Another axillary tail postbiopsy PEM image of the right axilla shows the carcinoma (star) and the FDG-avid axillary lymph node (arrow). In comparison to the whole-body positron emission tomography-computed tomography image [Figure 2], the lymph node has changed in size and configuration, indirectly indicating correct biopsy sampling

Click here to view
Figure 4: (a) Photograph of the ultrasound core needle biopsy specimen positioned on the detector plate of the PEM scanner. (b) Corresponding PEM image of the of the specimen reveals FDG activity within the core samples, confirming successful sampling of the lymph node

Click here to view



   Discussion Top


Our case is of a woman with right axillary invasive ductal carcinoma whose PET-CT scan showed FDG-avid right axillary lymphadenopathy. Whole-body PET-CT imaging was performed approximately 90 minutes after 18 F-FDG injection, followed by PEM imaging approximately 6 hours after injection. We additionally performed PEM imaging of the core-needle biopsy sample to assess for correct sampling.

The patient was injected with 362.6 MBq of 18 F-FDG at 07:45. PEM images of the right axilla were acquired at 13:47 and of the biopsy samples at 14:50. 18 F-FDG has a physical half-life of 110 min. 6 hours (3.3 half-lives) had elapsed between the initial administration of the radiotracer and the post-biopsy PEM images of the axilla leaving approximately 37 MBq of 18 F-FDG activity available for PEM imaging. The time difference between the initial administration of 18 F-FDG and PEM imaging of the core-needle biopsy sample was approximately 7 h or 3.8 half-lives, leaving approximately 25.9 MBq of 18 F-FDG activity available for PEM imaging.

Our case illustrates that activity as low as 25.9 MBq 18 F-FDG can be potentially utilized for PEM imaging, a level well below that of conventional recommendations of 185 to 370 MBq. Some recent studies have analyzed using lower levels of activity of 18 F-FDG for PEM imaging. MacDonald and colleagues. studied different activities of 18 F-FDG in phantom models in PEM imaging. They found that detection sensitivity was ≥90% for injected activities as low as 99.9 MBq for lesions ≥8 mm. [4] Mercier and colleagues compared the administration of 370 MBq and 185 MBq of 18 F-FDG for PEM imaging and found no differences between the two activities in quality scores and BI-RADS characterization of lesions. [5] A study by Kalinyak and Schilling simulated activity reduction for PEM by reducing the number of coincident event counts used to reconstruct images. [6] They found no significant change in the sensitivity for lesion detection when the coincident event counts were reduced to represent an activity of 185 MBq. [6] Our case indicates an even lower level of activity may be sufficient for adequate PEM imaging.

A unique advantage that PEM offers over breast MRI is the ability to image not only the biopsy site but also the biopsy specimen to confirm adequate sampling. Our post-biopsy PEM images revealed that our targeted right axillary lymph node was less conspicuous and had changed in shape after the biopsy in comparison to the pre-biopsy PET-CT, indirectly suggesting correct sampling of the lymph node. In addition, PEM imaging of the core-needle biopsy samples confirmed FDG activity within portions of the samples, indicating that the correct area was sampled. This potential use of PEM to assess the adequacy of the biopsy samples has been previously described in the literature with PEM-guided biopsy samples using vacuum-assisted devices. However, to our knowledge, 18 F-FDG activity in much smaller core-needle ultrasound biopsy samples has not been previously documented.

In summary, we discuss a case of utilization of lower levels of 18 F-FDG activity in PEM imaging to reduce radiation dose, while still maintaining the ability to adequately visualize the patient's known breast cancer in vivo as well as assessing the adequacy of non-vacuum-assisted core-needle biopsy sampling in vitro. Since PEM imaging is being increasingly utilized, future studies are needed to confirm the validity of using these lower levels of radiotracer activity for PEM imaging.

 
   References Top

1.
Berg WA, Madsen KS, Schilling K, Tartar M, Pisano ED, Larsen LH, et al. Breast cancer: Comparative effectiveness of positron emission mammography and MR imaging in presurgical planning for the ipsilateral breast. Radiology 2011;258:59-72.  Back to cited text no. 1
    
2.
Kalles V, Zografos GC, Provatopoulou X, Koulocheri D, Gounaris A. The current status of positron emission mammography in breast cancer diagnosis. Breast Cancer 2013;20:123-30.  Back to cited text no. 2
    
3.
Rosen EL, Eubank WB, Mankoff DA. FDG PET, PET/CT, and breast cancer imaging. Radiographics 2007;27 Suppl 1:S215-29.  Back to cited text no. 3
    
4.
Macdonald LR, Wang CL, Eissa M, Haseley D, Kelly MM, Liu F, et al. Positron emission mammography (PEM): Effect of activity concentration, object size, and object contrast on phantom lesion detection. Med Phys 2012;39:6499-508.  Back to cited text no. 4
    
5.
Mercier G, Vassilakis N, Brennan J, DeVita F, Kachnic L. Clinical performance of low activity positron emission mammography. Oncology: Basic, translational and therapy. J Nucl Med 2012;53:1231.  Back to cited text no. 5
    
6.
Kalinyak J, Schilling K. Radiotracer activity reduction in high resolution breast PET imaging. European Congress of Radiology; 2012. Available from: http://www.posterng.netkey.at/esr/viewing/index.php?module=viewing_poster and pi=112705. [Last accessed on 2014 Mar 30].  Back to cited text no. 6
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]


This article has been cited by
1 Ultrasound Imaging Technologies for Breast Cancer Detection and Management: A Review
Rongrong Guo,Guolan Lu,Binjie Qin,Baowei Fei
Ultrasound in Medicine & Biology. 2017;
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Case Report
   Discussion
    References
    Article Figures

 Article Access Statistics
    Viewed1345    
    Printed50    
    Emailed0    
    PDF Downloaded157    
    Comments [Add]    
    Cited by others 1    

Recommend this journal