ORIGINAL ARTICLE
Year : 2013  |  Volume : 12  |  Issue : 3  |  Page : 87-93

Assessment of Inter-modality Spatial Alignment Accuracy in Hybrid Single Photon Emission Computed Tomography in Patients with Hand and Wrist Pain


Department of Nuclear Medicine, Guy's and St. Thomas' Hospital NHS Foundation Trust, London, United Kingdom

Date of Web Publication14-Jul-2014

Correspondence Address:
Dr. Cheng Xie
14 Orion Building, 90 Navigation Street, Birmingham
United Kingdom
Login to access the Email id


DOI: 10.4103/1450-1147.136732

PMID: 25214811

Rights and Permissions
   Abstract 

Single photon emission computed tomography (SPECT) and computed tomography (CT) integrated in one system (SPECT/CT) is an effective co-registration technique that helps to localize and characterize lesions in the hand and wrist. However, patient motion may cause misalignment between the two modalities leading to potential misdiagnosis. The aim of the present study was to evaluate the hardware-based registration accuracy of multislice SPECT/CT of the hand and wrist and to determine the effect of misalignment errors on diagnostic accuracy. A total of 55 patients who had multislice SPECT/CT of the hand and wrist between July 2008 and January 2010 were included. Two reviewers independently evaluated the fused images for any misalignments with six degrees of freedom: Translation and rotation in the X, Y and Z directions. The results were tested against an automated fusion tool (Syntegra). More than half of the patients had moved during SPECT scanning (Reviewer 1: 29 patients; Reviewer 2: 30 patients) and they all originated in the Y-direction translation (vertical hand motion). Five fused images had significant misalignment errors that could have led to misdiagnosis. The Wilcoxon test indicated statistically non-significant difference (P > 0.05) between reviewers and statistically non-significant difference between the reviewers and software registration. The study also showed high inter-reviewer agreement (κ = 0.87). Hand movement during the SPECT scan was common, but significant misalignments and subsequent misdiagnosis were infrequent. Future studies should investigate the use of hand and wrist immobilization devices and reductions of scan time to minimize patient motion.

Keywords: Image fusion, registration, single photon emission computed tomography/computed tomography


How to cite this article:
Xie C, Gnanasegaran G, Mohan H, Livieratos L. Assessment of Inter-modality Spatial Alignment Accuracy in Hybrid Single Photon Emission Computed Tomography in Patients with Hand and Wrist Pain. World J Nucl Med 2013;12:87-93

How to cite this URL:
Xie C, Gnanasegaran G, Mohan H, Livieratos L. Assessment of Inter-modality Spatial Alignment Accuracy in Hybrid Single Photon Emission Computed Tomography in Patients with Hand and Wrist Pain. World J Nucl Med [serial online] 2013 [cited 2020 Sep 18];12:87-93. Available from: http://www.wjnm.org/text.asp?2013/12/3/87/136732


   Introduction Top


Integrated dual-modality single photon emission computed tomography/computed tomography (SPECT/CT) provides the means to combine imaging modalities for a direct functional and anatomical correlation to precisely localize and characterize abnormalities. [1],[2],[3],[4] Hybrid SPECT/CT imaging is particularly useful for assessing hand and wrist pain. [5] The anatomy of the hand especially the wrist joint consists of multiple articulating bones and its complex anatomy makes diagnosis and management of hand and wrist pain a challenge. [6] This is when radionuclide bone imaging is most valuable. Evidence showed that bone scans are highly sensitive in detecting orthopedic lesions. [7],[8] Bone scan alone is limited by its poor specificity and inferior image quality, it is often difficult to determine the precise location of the lesion amongst the close association of the carpal bones. [9] However, modern hybrid SPECT/CT imaging combines the advantages of both modalities - the underlying CT image allows clinicians to distinguish and evaluate each carpal bone individually, whereas the overlying SPECT image provides the exact location of the metabolic pathology. SPECT/CT has improved specificity and diagnostic confidence; it has shown to be a useful investigation in patients with hand and wrist pain. [10],[11],[12]

The co-registration of images from the hybrid SPECT/CT scanner relies on the common couch to maintain patient position throughout the scanning process. This technique reduces differences in patient positioning, thus limiting misregistration between the two modalities. However, experience from dual modality positron emission tomography (PET)/CT imaging has indicated that misregistration errors were still present in certain situations. [13],[14],[15] Studies have indicated that patient motion particularly during the long SPECT or PET scanning was the major contributor to the misregistration errors in hybrid systems. [16],[17],[18]

Over recent years, a series of publications have reported on limitations, artefacts and image distortions of SPECT and CT. [18],[19],[20],[21],[22],[23],[24],[25] Patient motion was identified as a major contributor to SPECT/CT misregistration. This emphasizes the importance of patient preparation-the need for patient comfort during scanning and reducing scan time to a minimum. Patient motion resulted in misregistration artefacts, most noticeably around the boundaries of the moved limb or organ.

At present, no data is available on the localization accuracy of hybrid SPECT/CT of the hand and wrist. This information is particularly important since a small misalignment between the SPECT and CT images could potentially position the hotspot onto a nearby carpal bone instead of the actual lesion. The result may be misleading and potentially delay appropriate treatment. Therefore, the aim of this study was to evaluate the registration accuracy of hybrid SPECT/CT imaging of the hand and wrist and determine the effect of the misalignment errors on the diagnostic accuracy.


   Materials and Methods Top


Patients

In this retrospective study, 55 consecutive patients (32 women and 23 men, 22-74 years of age, mean age of 45 years) who had 99mTc-MDP SPECT/CT of the hand and wrist between July 2008 and January 2010 in the Nuclear Medicine Department were included.

Imaging protocol

Dual-phase planar bone scans were performed, followed by SPECT/CT imaging. All the patients received intravenous injections of 750 MBq 99mTc-MDP. 5 min post-injection, blood pool images of the hands (positioned palms down on the camera face) were acquired using a 256 × 256 pixel matrix. At 3 h post-injection, delayed phase images of the hands were acquired in the same manner. SPECT/CT imaging followed straight after.

Image acquisition

For SPECT/CT imaging, patients were placed in the "Superman" position [Figure 1]. Patients lay prone with their head usually turned to one side allowing them to lie as flat as possible. Arms were straightened above the head toward the center of the couch. If only one hand was to be imaged, then only that arm was required above the head. This position limited CT irradiation within the area of interest and minimized irradiation of critical organs such as the head and the eyes. It also optimized the proximity of the gamma camera for better resolution. Pillows and foam blocks were placed under the patients' arms, head and lower legs for support and comfort. When the patient was made as comfortable as possible, their hands were taped palms down flat on a plastic board and finally ready for scan.
Figure 1: The "Superman" position for single photon emission computed tomography/computed tomography scanning

Click here to view


Scanning was performed on the dual-head gamma camera SPECT/CT system, the Philips Precedence 16 [Figure 2]. A CT scout view was first used to define the CT and SPECT fields. Next, localization CT was performed on the hands. The settings were 120 kV, 100 mAs/slice, a pitch of 1.188 with 0.75 s rotation time and collimation of 16 × 0.75 mm. The reconstructed images had 0.683 mm transverse pixel size and slice thickness of 1.5 mm. The CT scan took about 1 min to complete.
Figure 2: The dual-head gamma camera single photon emission computed tomography/computed tomography system - Philips Precedence 16

Click here to view


For SPECT, the emitted gamma rays from the hand were acquired over 360°. Low energy high resolution, parallel hole collimators were used. A total of 128 projections and each with a duration of 20 s. Data were acquired into a 128 × 128 matrix, which resulted in a pixel size of 4.664 mm. The SPECT scan took approximately 25 min.

Image analysis

The raw SPECT data were reconstructed on a Jet Stream Workspace (v3.0, Philips Medical Systems, USA) with filtered back projection and Butterworth filter with cut-off of 0.4 cycle/pixel and order of four. Reconstructed transverse slices were sent to a Philips Extended Brilliance Workstation (EBW) for review of the images fused with the co-registered CT images. The process of defining the co-registration spatial transformation parameters involved scanning a phantom that consists of six point sources visible in both SPECT and CT. The algorithm defined the center of mass for each point source and the spatial transformation parameters that will result in the centroids overlapping on the two modalities. These parameters are saved on the system and consequently applied to all patient data.

Two experienced consultants in Nuclear Medicine were involved in image analysis. They independently evaluated the fused images on EBW using the manufacturer's fusion display tool (Fusion Application Suite V1.2E). This tool provided the reviewers with transverse, coronal and sagittal planes of the fused images for assessment. In each plane, the reviewers went through all the fused slices and if necessary adjusted the linear grey scale and color fusion display to identify anatomical landmarks to help determine the accuracy of alignment. When a reviewer identified a misalignment between the SPECT and CT images, they were able to use the display tool to re-align the images manually across six degrees of freedom: Translation in the X, Y and Z directions, rotation in the X, Y and Z directions [Figure 3]a and b. Any misalignment resulting in spatial displacement of more than 4.66 mm (one pixel size) was defined as a significant misalignment. Misalignment of less than one pixel size was difficult to correct for given the spatial resolution of the SPECT modality and hence regarded as non-significant. In addition to manual re-alignment, the images were analyzed by an automated fusion tool (Syntegra version 2.2). Syntegra was included in the Philips JETStream platform and it performed rigid-body co-registration with a normalized mutual information algorithm. The spatial transformations required to correct the misalignments were recorded for statistical analysis.
Figure 3:

Click here to view


To illustrate the effect that inter-modality misalignment may have had on diagnostic accuracy, the reviewers further evaluated the images that had significant misalignments. They compared fused images before and after misalignment correction to see whether the deviation of the overlying hotspot on the CT image may have resulted in a misdiagnosis.

Statistical analysis

All statistical analysis was carried out on Statistical Package for the Social Sciences (SPSS, v17). The level of significance was set at P < 0.05. The Wilcoxon test was used to evaluate the inter-observer reproducibility of this method. This test allowed direct comparison of the two reviewers' misalignment corrections in each of the misalignment directions. Furthermore, the Kappa measurement of agreement was used to assess the reviewers' agreement of the misalignments.

Manual misalignment corrections performed by the reviewers were also compared with automated correction carried out by the Syntegra software. The Wilcoxon test was used to compare each reviewer's results with those from Syntegra.


   Results Top


Misalignment analysis

Evaluation of the SPECT/CT fused images revealed that more than half of the patients had moved during SPECT scanning (Reviewer 1: 52%; Reviewer 2: 54%). Most of the misalignments observed were translations along the Y-direction (vertical hand movement) [Table 1] and were evident in the transverse and sagittal planes [Figure 3]a. This indicated that the patients' hands were moving mainly up/down than any other direction during the SPECT scan.
Table 1: The number of patients with misalignment identified by the reviewers

Click here to view


Furthermore, non-rigid body movement was also observed: There were six cases of thumb movement and one case of finger movement [Figure 4]. In these cases, other parts of the hand and wrist were aligned.
Figure 4: Left: The patient's thumb moved during the single photon emission computed tomography (SPECT) scan while all other parts were aligned. Right: The patient's third, fourth and fifth digits moved during the SPECT scan. The misalignment is most visible on the fifth digit

Click here to view


Diagnostic accuracy analysis

Reviewers identified five patients with significant vertical hand motion. The subsequent misalignments of the fused imaged could have led to misdiagnosis if manual misalignment corrections were not performed. [Table 2] shows the location of the increased focal uptake of the tracer before and after misalignment correction for these five patients. [Figure 5]a and b are images of a patient to demonstrate the change in location of the focal tracer uptake after misalignment correction, which could have led to misdiagnosis if corrections did not take place.
Figure 5:

Click here to view


Inter-reviewer agreement

The Wilcoxon test directly compared the two reviewers' misalignment corrections in all six directions and found there was a statistically non-significant difference (P > 0.05) in misalignment measurements in all six directions between the two reviewers. The Kappa value also demonstrated high inter-reviewer agreement (0.87).

The Wilcoxon test was used to directly test the reviewers' manual misalignment corrections against the automated corrections carried out by Syntegra. The results indicated there was a statistically non-significant difference (P > 0.05) between each of the reviewers and Syntegra in correcting for misalignment errors.


   Discussion Top


Recent advances in medical imaging technology have led to the development the hybrid SPECT/CT with multi-detector CT components of fully diagnostic capabilities. This allows fast and effective co-registration of functional and anatomical information. The registration accuracy is particularly important for a confident diagnosis in the hand and wrist due to their complex anatomy. This study was designed to assess the registration accuracy of hybrid SPECT/CT imaging of the hand and wrist and determine the effect of the misalignment errors on the diagnostic accuracy. Our results indicated that approximately half (54%) of the 55 patients had moved between the CT and SPECT scan as determined by the reviewers' reports. Most of the misalignments were translations in the Y-direction. This suggested that the patients' hands were moving mainly up or down than any other direction during the SPECT scan.

Even though more than half of the patients had moved during the SPECT scan or between the CT and SPECT scans, the majority of the subsequent misalignment errors did not interfere with diagnosis. This suggested that hybrid SPECT/CT can be considered to have a good degree of registration accuracy. Among the limited number of published works on this area, the study conducted by Nömayr et al. showed agreement with our study findings. Nömayr et al. evaluated the anatomical accuracy of hybrid SPECT/CT in 22 patient scans of the lower spine. Two radiologists and an automated fusion tool assessed the images. They reported anatomical inaccuracies below the single pixel width of 4.6 mm and concluded that the hardware approach of SPECT and CT image fusion was "nearly perfect." [26]

Hybrid SPECT/CT may be considered to have a good degree of registration accuracy, but the fact that more than half of the patients in our study did move during scanning and nearly one in five of the patients did produce a significant misalignment suggested that there were factors interfering with the registration accuracy.

One major factor to consider was the long scanning time, which compromised the positioning of the patient during the scan. In this study, the CT scan lasted only about a minute, but the SPECT scan took approximately 25 min to complete. During this long acquisition time, the patient was asked to lie still in the "superman" position. The procedure was made as comfortable as possible with the use of pillows and foam blocks. Yet it was likely that the patient experienced some discomfort or anxiety during the scan and triggered a slight hand movement. In addition, the patients may be already experiencing some pain or discomfort from the hand or wrist injury, which prevented them from lying still for long.

An equally important factor to consider was the immobilization procedures and devices used during scanning. In this study, the protocol was to use tape to fix the patient's hand palm down, flat on a plastic board. First, the tape was effective in fixing the hand and wrist to the plastic board, but the individual fingers and thumb were poorly immobilized onto the board. There was plenty of free space in between the thumb and fingers. This may have contributed to the occasional thumb or finger movements that occurred in this study [Figure 4]. Second, the amount of tape used and the position of the tape on the hand and wrist were variable in between radiographers and these factors could have contributed to the different degree of immobilization.

In this study, we observed that patients' hands were moving mainly along the vertical direction than any other direction during the SPECT scan. Two important contributing factors were identified. The first factor was the immobilization technique used to rest the patient's hand. This has to be firm enough to restrict rigid and affine movement of the object especially under the complexity of the possible degrees of freedom involved in the hand and wrist. According to the radiographers, they have observed that patients can be anxious and stressed during the beginning of the scan (second contributing factor) and sometimes they tense their hand and unknowingly pressed down hard on the board. As the scan progressed, the patients usually relax and their hands on the board sprung back up.

One factor that could have influenced the misalignments was the injury. The severity and timing of the hand or wrist injury could have inflicted more discomfort during the scan and resulted in more movement. We did not explore this relationship, but it would be interesting to examine in our future study.

Clinical relevance

We report that after misalignment correction, 5/55 fused images had demonstrated a change in location of the focal tracer uptake and could have led to misdiagnosis. This indicated that for every 100 patients who had undergone hybrid SPECT/CT scan of the hand and wrist, there is the potential of approximately ten misdiagnosed cases if manual misalignment corrections had not taken place.

Solutions and future work

In this study, we have identified several causes to patient motion and subsequent misalignment errors. We intend to address these in our future work.

The first practical solution is to implement improved patient immobilization techniques such as the use of a thermoplastic mesh [Figure 6] to immobilize the hand and wrist. The thermoplastic mesh, once submerged in hot water, can be easily molded to fit around each patient's hand. Once cooled, it can immobilize the hand and wrist as well as the individual fingers and thumb. The mesh can be re-used and re-molded for the next patient. It is made from lightweight plastic and will not affect the SPECT or CT images via photon attenuation. We intend to implement the thermoplastic mesh in a prospective study on a sample of patients and assess the registration outcomes.
Figure 6: Thermoplastic mesh made from lightweight plastic. It can be individually molded to fit the patient's hand providing better support and immobilization

Click here to view


One of the main causes of patient motion was the duration of the SPECT scan and this is the most challenging obstacle. A trade-off exists between scan time and SPECT image quality. Reducing the scan time would reduce the probability for patient motion, thus minimizing misalignment errors. On the other hand, a reduction in acquisition time would reduce the number of acquired counts and thus increase noise levels. Several published reports on advance image reconstruction methods, such as those incorporating depth-dependent resolution models, have demonstrated promising results in reducing SPECT scan time in myocardial perfusion studies by a half to three-quarter of the original time [27],[28],[29] without compromising image quality or diagnostic accuracy. Thus application of such reconstruction schemes would also be relevant in the context of containing misalignment artefacts due to patient motion.

Study limitations

In this study two reviewers manually assessed all the SPECT/CT images. There was the possibility of subjective errors and inconsistency in between reviewers, which meant that the reproducibility and accuracy of this study method were not absolute. However, results from the Wilcoxon test and the Kappa measurement of agreement did demonstrate a high level of agreement and reproducibility between the two reviewers. Furthermore, there was a statistically non-significant difference when each of the reviewers' results was compared against the automated fusion tool Syntegra when applied to cases of significant misalignments. Therefore, given the subjective nature of observer-based studies the agreement in this case was satisfactory.

Another potential drawback of this study lies within this comparison between the reviewers' manually assessed misalignments and Syntegra's automated fusion results. This comparison was only limited to the images that the reviewers had identified to possess significant misalignments. Only these images were subjected to further analysis on Syntegra. The rest of the images were not processed on Syntegra due to the time consuming nature of performing automated image fusion. As a result, we were unable to assess the rest of the manually corrected images against the automated corrected images. The reliability of this study method would have been more accurately assessed if all 55 SPECT/CT images were analyzed by Syntegra and compared with each reviewer's results.


   Conclusion Top


A dual-modality integrated hybrid SPECT/CT scanner is an efficient and effective co-registration technique that provides direct functional and anatomical correlation to localize and characterize the abnormality in the hand and wrist. Hand movement during the SPECT scan is common, but the degree of movements leading to significant misalignments and subsequently affecting diagnostic accuracy are infrequent. The registration accuracy depends on patient compliance to maintain still during the scan; hence patient comfort during scanning is the prerequisite of an ideal scan.

 
   References Top

1.von Schulthess GK, Pelc NJ. Integrated-modality imaging: The best of both worlds. Acad Radiol 2002;9:1241-4.  Back to cited text no. 1
[PUBMED]    
2.Schillaci O. Hybrid SPECT/CT: A new era for SPECT imaging? Eur J Nucl Med Mol Imaging 2005;32:521-4.  Back to cited text no. 2
[PUBMED]    
3.Townsend DW, Cherry SR. Combining anatomy and function: The path to true image fusion. Eur Radiol 2001;11:1968-74.  Back to cited text no. 3
    
4.Even-Sapir E, Keidar Z, Bar-Shalom R. Hybrid imaging (SPECT/CT and PET/CT) - Improving the diagnostic accuracy of functional/metabolic and anatomic imaging. Semin Nucl Med 2009;39:264-75.  Back to cited text no. 4
    
5.Fogelman I. Bone scanning. In: Maisey M, Britton K, Gilday D, editors. Clinical Nuclear Medicine. Cambridge: Chapman & Hall; 1991. p. 131-57.  Back to cited text no. 5
    
6.Forman TA, Forman SK, Rose NE. A clinical approach to diagnosing wrist pain. Am Fam Physician 2005;72:1753-8.  Back to cited text no. 6
    
7.Savelli G, Maffioli L, Maccauro M, De Deckere E, Bombardieri E. Bone scintigraphy and the added value of SPECT (single photon emission tomography) in detecting skeletal lesions. Q J Nucl Med 2001;45:27-37.  Back to cited text no. 7
    
8.Prather JL, Nusynowitz ML, Snowdy HA, Hughes AD, McCartney WH, Bagg RJ. Scintigraphic findings in stress fractures. J Bone Joint Surg Am 1977;59:869-74.  Back to cited text no. 8
[PUBMED]    
9.Even-Sapir E. Imaging of malignant bone involvement by morphologic, scintigraphic, and hybrid modalities. J Nucl Med 2005;46:1356-67.  Back to cited text no. 9
[PUBMED]    
10.Papathanassiou D, Bruna-Muraille C, Jouannaud C, Gagneux-Lemoussu L, Eschard JP, Liehn JC. Single-photon emission computed tomography combined with computed tomography (SPECT/CT) in bone diseases. Joint Bone Spine 2009;76:474-80.  Back to cited text no. 10
    
11.Gnanasegaran G, Barwick T, Adamson K, Mohan H, Sharp D, Fogelman I. Multislice SPECT/CT in benign and malignant bone disease: When the ordinary turns into the extraordinary. Semin Nucl Med 2009;39:431-42.  Back to cited text no. 11
    
12.Huellner MW, Bürkert A, Schleich FS, Schürch M, Hug U, von Wartburg U, et al. SPECT/CT versus MRI in patients with nonspecific pain of the hand and wrist - A pilot study. Eur J Nucl Med Mol Imaging 2012;39:750-9.  Back to cited text no. 12
    
13.Nakamoto Y, Tatsumi M, Cohade C, Osman M, Marshall LT, Wahl RL. Accuracy of image fusion of normal upper abdominal organs visualized with PET/CT. Eur J Nucl Med Mol Imaging 2003;30:597-602.  Back to cited text no. 13
    
14.Zaki M, Suga K, Kawakami Y, Yamashita T, Shimizu K, Seto A, et al. Preferential location of acute pulmonary thromboembolism induced consolidative opacities: Assessment with respiratory gated perfusion SPECT-CT fusion images. Nucl Med Commun 2005;26:465-74.  Back to cited text no. 14
    
15.Cohade C, Osman M, Marshall LN, Wahl RN. PET-CT: Accuracy of PET and CT spatial registration of lung lesions. Eur J Nucl Med Mol Imaging 2003;30:721-6.  Back to cited text no. 15
    
16.Chen J, Caputlu-Wilson SF, Shi H, Galt JR, Faber TL, Garcia EV. Automated quality control of emission-transmission misalignment for attenuation correction in myocardial perfusion imaging with SPECT-CT systems. J Nucl Cardiol 2006;13:43-9.  Back to cited text no. 16
    
17.Blodgett TM, McCook BM, Federle MP. Positron emission tomography/computed tomography: Protocol issues and options. Semin Nucl Med 2006;36:157-68.  Back to cited text no. 17
    
18.Gnanasegaran G, Cook G, Adamson K, Fogelman I. Patterns, variants, artifacts, and pitfalls in conventional radionuclide bone imaging and SPECT/CT. Semin Nucl Med 2009;39:380-95.  Back to cited text no. 18
    
19.Forstrom LA, Dunn WL, O'Connor MK, Decklever TD, Hardyman TJ, Howarth DM. Technical pitfalls in image acquisition, processing, and display. Semin Nucl Med 1996;26:278-94.  Back to cited text no. 19
    
20.Howarth DM, Forstrom LA, O'Connor MK, Thomas PA, Cardew AP. Patient-related pitfalls and artifacts in nuclear medicine imaging. Semin Nucl Med 1996;26:295-307.  Back to cited text no. 20
    
21.O'Connor MK. Instrument- and computer-related problems and artifacts in nuclear medicine. Semin Nucl Med 1996;26:256-77.  Back to cited text no. 21
[PUBMED]    
22.Ogawa K. Image distortion and correction in single photon emission CT. Ann Nucl Med 2004;18:171-85.  Back to cited text no. 22
[PUBMED]    
23.Popilock R, Sandrasagaren K, Harris L, Kaser KA. CT artifact recognition for the nuclear technologist. J Nucl Med Technol 2008;36:79-81.  Back to cited text no. 23
    
24.Wilting JE, Timmer J. Artefacts in spiral-CT images and their relation to pitch and subject morphology. Eur Radiol 1999;9:316-22.  Back to cited text no. 24
    
25.Fleischmann D, Rubin GD, Paik DS, Yen SY, Hilfiker PR, Beaulieu CF, et al. Stair-step artifacts with single versus multiple detector-row helical CT. Radiology 2000;216:185-96.  Back to cited text no. 25
    
26.Nömayr A, Römer W, Strobel D, Bautz W, Kuwert T. Anatomical accuracy of hybrid SPECT/spiral CT in the lower spine. Nucl Med Commun 2006;27:521-8.  Back to cited text no. 26
    
27.Ali I, Ruddy TD, Almgrahi A, Anstett FG, Wells RG. Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med 2009;50:554-62.  Back to cited text no. 27
    
28.Venero CV, Heller GV, Bateman TM, McGhie AI, Ahlberg AW, Katten D, et al. A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol 2009;16:714-25.  Back to cited text no. 28
    
29.DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction "quarter-time" gated myocardial perfusion SPECT functional imaging: A comparison to "full-time" ordered subset expectation maximum. J Nucl Cardiol 2009;16:736-52.  Back to cited text no. 29
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
 
 
    Tables

  [Table 1], [Table 2]


This article has been cited by
1 Technical Pitfalls and Limitations of SPECT/CT
Lefteris Livieratos
Seminars in Nuclear Medicine. 2015; 45(6): 530
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1926    
    Printed61    
    Emailed0    
    PDF Downloaded232    
    Comments [Add]    
    Cited by others 1    

Recommend this journal